Testo 420
- Description
Description
- Differential pressure measurement, useful for checking filters in air conditioning and ventilation systems  Calculate flow velocity and volume flow using an optional pitot tube  Save measurement data for analysis and documentation  Match with a Testo CEC compliant air flow capture hood to measure volume flows at supplies and returns (0563 4200)
- Differential Pressure:
- Measuring range -0.480 to 0.480 InHâ‚‚O / -120 to 120 Pa 19.7 to 2756 fpm / 0.1 to 14 m/s
- Accuracy ±2 % of mv 0.002 InH₂O at 72 °F, 405 InH₂O / ±2 % of mv 0.5 Pa at 22 °C, 1013 hPa
- Resolution 0.001 InHâ‚‚O / 0.001 Pa 0.01 fpm / 0.01 m/s
- Reaction time 1 sec (t90)
>
- Absolute Pressure
- Measuring range 280 to 440 InHâ‚‚O / 700 to 1100 hPa
- Accuracy ±1 InH₂O / ±3 hPa
- Resolution 0.1 InHâ‚‚O / 0.1 hPa
- Reaction time 1 sec
>
- Dimensions 6 x 3 x 1 in. / 150 x 85 x 35 mm
- Operating temperature 23° to 122 °F / -5 to 50 °C
- EU-/EG-guidelines EU-guideline 2014/30/EU
- Battery type Alkaline, type AA
- Battery life 40 h
- Display size 3.5
- Display function Dot matrix
- Display ligthing with illumination
- Interface Micro-USB
- Memory 2 GB (internal)
- Storage temperature -4° to 140 °F / -20 to 60 °C
>
Each Unit Includes
- Testo 420 differential pressure measuring instrument
- Batteries
- Certificate of calibration