Fluke 725-L VIP Kit

Fluke 725-L VIP Kit

0 out of 5

$1,667.00

SKU: 725E-L-VIP Category: Tag:
View cart

Description

Exclusive VIP Kit includes FREE Universal Carrying Case, FREE Mini Plug To Pigtail T/C Lead Set , FREE Toolpack , and FREE Pack of Green Calibration Labels . The one multifunction calibrator you need for all of your process calibrations.

  • Value Incentive Package
  • LIMITED TIME ONLY! Hurry Offer Ends Soon!
  • Temperature Calibrator (745), with FREE Carrying Case (C-101), & FREE T/C Lead Set (22171T), & FREE Toolpak Hanger (TPAK80-4) FREE PST Tool (61010101), & FREE Calibration Labels
  • Over $150 Savings
  • 725 Can source/simulate volts, mA, RTDs, thermocouples, frequency, and ohms. Measure and source pressure using Fluke 700P series pressure modules. Has Dual display shows input and output. Dual display shows input and output. Programmable step and ramp for automated testing. Powered by 4 AA batteries (25 hours typical operation)
  • FREE C-101 Universal Carrying Case! Zippered section holds Multimeters, Single Function Calibrators, etc. Velcro pouch for leads and accessories. Snappable belt loops on back hold up to a 4 wide belt.
  • FREE 22171T- JKTE Mini Plug To Pigtail T/C Lead Set
  • FREE TPAK80-4 Toolpack -Hang your meter by a magnet or a strap hanger. Locks securely to the meter. Includes 6 strap, 17 strap, and slip on magnet.
  • FREE PST Tool- PST Consists of the following all in one! Needlenose and Regular Pliers, Wire Cutters, Hard-Wire Cutters, Clip-Point Knife, Metal/Wood File, 8in/20 cm Ruler, Can/Bottle Opener, [Small, Medium, Large and Phillips] Screwdriver, Awl Punch
  • FREE Pack of Green Calibration Labels

>

Each Unit Includes

  • Fluke 725
  • TL75 Test Leads
  • AC70A Test Clips
  • One pair of stackable test leads
  • Product Overview Manual in 14 different languages
  • User Manual on CD-Rom
  • TPAK80-4 Took Pack Hanger
  • C-101 Carrying Case
  • 22171T J,K,T,E Mini T/C Lead Set
  • Leatherman PST Tool
  • Calibration Labels

>

Measurement Accuracy
Voltage DC 30.000 V 0.02% 2 counts (upper display)
30.000 V 0.02% 2 counts (lower display)
100.00 mV 0.02% 2 counts
-10.00 mV to 75.00 mV 0.025% 1 count (via TC connector)
Current DC 24.000 mA 0.02% 2 counts
Resistance 0.0 to 400.0 Ω 0.1 Ω (4-wire), 0.15 Ω (2- and 3-wire)
401 to 1500 Ω 0.5 Ω (4-wire), 1 Ω (2- and 3-wire)
1500 to 3200 Ω 1 Ω (4-wire), 1.5 Ω (2- and 3-wire)
Frequency 2.0 to 1000.0 CPM 0.05% 1 count
1.0 to 1100.0 Hz 0.05% 1 count
1.00 to 10.00 kHz 0.05% 1 count
Sensitivity 1 V peak-to-peak-minimum
Pressure Accuracy from 0.025% of range using any of 29 pressure modules (for detailed specifications refer to pressure modules in options and accessories). Modules available for differential, gauge, vacuum, absolute, dual and high pressure.
Source Accuracy
Voltage DC 100.00 mV 0.02% 2 counts
10.000 V 0.02% 2 counts
-10.00 mV to 75.00 mV 0.025% 1 count (via TC connector)
Current DC 24.000 mA (source) 0.02% 2 counts
24.000 mA (simulate) 0.02% 2 counts
Resistance 15.0 to 400.0 Ω 0.15 Ω (exc. current 0.15 to 0.5 mA), 0.1 Ω (exc. current 0.5 to 2 mA)
401 to 1500 Ω 0.5 Ω (excitation current 0.05 to 0.8 mA)
1500 to 3200 Ω 1 Ω (excitation current 0.05 to 0.4 mA)
Frequency 2.0 to 1000.0 CPM 0.05%
1.0 to 1100.0 Hz 0.05%
1.00 to 10.00 kHz 0.25%
Waveform 5 V p-p squarewave, -0.1 V offset
RTDs and Thermocouples
Measure accuracy NI-120 0.2°C
PT-100 (385) 0.33°C
PT-100 (393) 0.3°C
PT-100 (JIS) 0.3°C
PT-200 (385) 0.2°C
PT-500 (385) 0.3°C
PT-1000 (385) 0.2°C
Resolution 0.1°C
J 0.7°C
K 0.8°C
T 0.8°C
E 0.7°C
R 1.8°C
S 1.5°C
B 1.4°C
L 0.7°C
U 0.75°C
N 0.9°C
Resolution J, K, T, E, L, N, U: 0.1°C, 0.1°F B, R, S: 1°C, 1°F
XK 0.6°C
BP 1.2°C
Source accuracy NI-120 0.2°C
PT-100 (385) 0.33°C
PT-100 (393) 0.3°C
PT-100 (JIS) 0.3°C
PT-200 (385) 0.2°C
PT-500 (385) 0.3°C
PT-1000 (385) 0.2°C
Resolution 0.1°C
Accuracy stated for 4-wire measurement.
J 0.7°C
K 0.8°C
T 0.8°C
E 0.7°C
R 1.4°C
S 1.5°C
B 1.4°C
L 0.7°C
U 0.75°C
N 0.9°C
Resolution J, K, T, E, L, N, U: 0.1°C, B, R, S: 1°C
XK 0.6°C
BP 1.2°C
Specifications
Ramp functions Source functions Voltage, current, resistance, frequency, temperature
Ramps Slow ramp, Fast ramp, 25% step-ramp
Loop power function Voltage 24 V
Accuracy 10%
Maximum current 22 mA, short circuit protected
Step functions Source functions Voltage, current, resistance, frequency, temperature
Steps 25% of range, 100% of range
Environmental Specifications
Operating temperature -10°C to 55°C
Storage temperature -20°C to 71°C
Operating altitude 3000 m
Safety Specifications
Agency approvals EN 61010-1:1993, ANSI/ISA S82.01-1994, CAN/CSA C22.2 No 1010.1:1992
Mechanical and General Specifications
Size 130 x 236 x 61 mm
Weight 0.65 kg
Batteries 4 AA alkaline batteries
Battery replacement Separate battery compartment, accessable without breaking calibration seal
Side port connections Pressure module connector, also used for remote real-time programming